%0 Journal Article
%A Wang, Yuting
%A May, Markus W
%A Gratz, Marcel
%A Ladd, Mark
%A Orzada, Stephan
%T Ultimate Intrinsic SNR in the Torso of Realistic Body Models.
%J Magnetic resonance in medicine
%V 95
%N 4
%@ 1522-2594
%C New York, NY [u.a.]
%I Wiley-Liss
%M DKFZ-2025-02647
%P 2409-2419
%D 2026
%Z #EA:E020#LA:E020# / 2026 Apr;95(4):2409-2419
%X This work aims to investigate how the ultimate intrinsic signal-to-noise ratio (uiSNR) varies with increasing static magnetic field B 0
%X <br clear="all" /><table border="0" width="100%"><tr><td>
%X <table align="center" cellspacing="0" cellpadding="2"><tr><td nowrap="nowrap" align="center">
%X B<sub>0</sub> </td></tr></table>
%X </td></tr></table>
%X
%X in the torso of realistic body models.A dipole cloud was positioned around the realistic body model and randomly excited. The volume integral solver MARIE was used to calculate the corresponding electromagnetic fields. The uiSNR maps were calculated using these electromagnetic bases and were fitted with the power law for different B 0
%X <br clear="all" /><table border="0" width="100%"><tr><td>
%X <table align="center" cellspacing="0" cellpadding="2"><tr><td nowrap="nowrap" align="center">
%X B<sub>0</sub> </td></tr></table>
%X </td></tr></table>
%X
%X ranges.The uiSNR could be reliably calculated in regions deeper than 3 cm, where convergence of uiSNR over the number of basis vectors was achieved. In a lower magnetic field range (from 0.55 to 3 T), the uiSNR increases roughly linearly versus B 0
%X <br clear="all" /><table border="0" width="100%"><tr><td>
%X <table align="center" cellspacing="0" cellpadding="2"><tr><td nowrap="nowrap" align="center">
%X B<sub>0</sub> </td></tr></table>
%X </td></tr></table>
%X
%X with small variation throughout the torso (Ella: uiSNR ∝ B0 0.96±0.07, Duke: uiSNR ∝ B0 0.98±0.10). In an upper magnetic field range (from 5 to 14 T), the uiSNR increases superlinearly in the torso (Ella: uiSNR ∝ B0 1.86±0.25, Duke: uiSNR ∝ B0 1.99±0.28), with a larger variation correlated to the heterogeneous structure of the body model.The superlinear scaling exponent in the upper magnetic field range indicates the promise of applying UHF MRI for body imaging.
%K MRI (Other)
%K in silico (Other)
%K torso (Other)
%K uiSNR (Other)
%F PUB:(DE-HGF)16
%9 Journal Article
%$ pmid:41299668
%R 10.1002/mrm.70202
%U https://inrepo02.dkfz.de/record/306610