| Home > Publications database > Absorbed dose in ion beams: comparison of ionisation- and fluence-based measurements. |
| Journal Article | DKFZ-2017-00656 |
; ; ; ; ;
2014
Oxford Univ. Press
Ashford
This record in other databases:

Please use a persistent id in citations: doi:10.1093/rpd/ncu004
Abstract: A direct comparison measurement of fluorescent nuclear track detectors (FNTDs) and a thimble ionisation chamber is presented. Irradiations were performed using monoenergetic protons (142.66 MeV, ϕ=3×10(6) cm(-2)) and carbon ions (270.55 MeV u(-1), ϕ=3 × 10(6) cm(-2)). It was found that absorbed dose to water values as determined by fluence measurements using FNTDs are, in case of protons, in good agreement (2.4 %) with ionisation chamber measurements, if slower protons and Helium secondaries were accounted for by an effective stopping power. For carbon, however, a significant discrepancy of 4.5 % was seen, which could not be explained by fragmentation, uncertainties or experimental design. The results rather suggest a W-value of 32.10 eV ± 2.6 %. Additionally, the abundance of secondary protons expected from Monte-Carlo transport simulation was not observed.
Keyword(s): Ions ; Protons ; Water ; Helium ; Carbon
|