Performance and safety assessment of an integrated transmit array for body imaging at 7 T under consideration of specific absorption rate, tissue temperature, and thermal dose.

 ;  ;  ;  ;  ;  ;  ;  ;  ;

2022
Wiley New York, NY

NMR in biomedicine 35(5), e4656 () [10.1002/nbm.4656]
 GO

This record in other databases:  

Please use a persistent id in citations: doi:

Abstract: In this study, the performance of an integrated body-imaging array for 7 T with 32 radiofrequency (RF) channels under consideration of local specific absorption rate (SAR), tissue temperature, and thermal dose limits was evaluated and the imaging performance was compared with a clinical 3 T body coil. Thirty-two transmit elements were placed in three rings between the bore liner and RF shield of the gradient coil. Slice-selective RF pulse optimizations for B1 shimming and spokes were performed for differently oriented slices in the body under consideration of realistic constraints for power and local SAR. To improve the B1 + homogeneity, safety assessments based on temperature and thermal dose were performed to possibly allow for higher input power for the pulse optimization than permissible with SAR limits. The results showed that using two spokes, the 7 T array outperformed the 3 T birdcage in all the considered regions of interest. However, a significantly higher SAR or lower duty cycle at 7 T is necessary in some cases to achieve similar B1 + homogeneity as at 3 T. The homogeneity in up to 50 cm-long coronal slices can particularly benefit from the high RF shim performance provided by the 32 RF channels. The thermal dose approach increases the allowable input power and the corresponding local SAR, in one example up to 100 W/kg, without limiting the exposure time necessary for an MR examination. In conclusion, the integrated antenna array at 7 T enables a clinical workflow for body imaging and comparable imaging performance to a conventional 3 T clinical body coil.

Keyword(s): body imaging at 7 T MRI ; thermal dose ; tissue temperature ; transmit antenna arrays

Classification:

Note: #EA:E020#LA:E020# / 2022 May;35(5):e4656

Contributing Institute(s):
  1. E020 Med. Physik in der Radiologie (E020)
Research Program(s):
  1. 315 - Bildgebung und Radioonkologie (POF4-315) (POF4-315)

Appears in the scientific report 2021
Database coverage:
Medline ; Clarivate Analytics Master Journal List ; Current Contents - Life Sciences ; DEAL Wiley ; Essential Science Indicators ; IF < 5 ; JCR ; NationallizenzNationallizenz ; SCOPUS ; Science Citation Index Expanded ; Web of Science Core Collection
Click to display QR Code for this record

The record appears in these collections:
Document types > Articles > Journal Article
Institute Collections > E020
Public records
Publications database

 Record created 2021-12-29, last modified 2024-02-29



Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)