TY - JOUR
AU - Lou, Feiyang
AU - Duan, Caohui
AU - Quan, Zhiyan
AU - Qu, Jianxun
AU - Guo, Chenlu
AU - Nagel, Armin M
AU - Lou, Xin
AU - Zhang, Xiaotong
TI - A Double-Layer 1H/23Na Transceiver Array for Human Brain MRI at 7T.
JO - NMR in biomedicine
VL - 38
IS - 12
SN - 0952-3480
CY - New York, NY
PB - Wiley
M1 - DKFZ-2025-02151
SP - e70162
PY - 2025
AB - With the increasing adoption of ultra-high-field (UHF) systems, sodium (23Na) imaging is undergoing clinical translation. However, existing commercially available dual-tuned coils, while demonstrating adequate sodium signal-to-noise ratio (SNR), fail to meet the requirements for high-resolution proton (1H) structural imaging, resulting in suboptimal workflow efficiency during integrated multi-nucleus examinations. In this work, we present a novel double-layer 1H/23Na transceiver array with eight channels per nucleus for human brain MRI at 7T. The proposed array achieved 64
KW - Humans
KW - Magnetic Resonance Imaging: instrumentation
KW - Brain: diagnostic imaging
KW - Sodium
KW - Signal-To-Noise Ratio
KW - Protons
KW - Equipment Design
KW - Sodium Isotopes
KW - RF coil (Other)
KW - sodium MRI (Other)
KW - ultra‐high‐field MRI (Other)
KW - Sodium (NLM Chemicals)
KW - Protons (NLM Chemicals)
KW - Sodium Isotopes (NLM Chemicals)
LB - PUB:(DE-HGF)16
C6 - pmid:41098051
DO - DOI:10.1002/nbm.70162
UR - https://inrepo02.dkfz.de/record/305422
ER -