Diagnostic performance of clinical photon-counting CT in the assessment of peri-implantitis - an ex vivo diagnostic accuracy study.

 ;  ;  ;  ;  ;  ;  ;  ;  ;  ;

2026
Springer Heidelberg

International journal of implant dentistry 12(1), 59 () [10.1186/s40729-026-00722-1]
 GO

Abstract: To evaluate the diagnostic performance of photon-counting computed tomography (PCCT) for the assessment of peri-implant bone defects and to compare it with cone-beam computed tomography (CBCT) at comparable radiation doses.Circular peri-implant defects were artificially created around 27 titanium implants placed in nine porcine mandibles. Defect morphology was digitized using an intraoral scanner (STL, stereolithography file format) as reference standard. Specimens were imaged with high-resolution CBCT (HD-CBCT) and PCCT at clinically commonly used dose levels, which were nearly identical for both modalities to ensure comparability. Vertical defect depth and horizontal defect width were measured at predefined locations by two calibrated readers. Agreement with the reference standard was assessed using paired statistical testing, Bland-Altman analysis, mean absolute error (MAE), relative error metrics, and intraclass correlation coefficients (ICC).For vertical defect depth (mean reference 4.68 ± 2.20 mm), PCCT demonstrated significantly lower MAE than CBCT (0.14 mm vs. 0.67 mm; p < 0.001) and narrower limits of agreement. Relative absolute error was 4.3% for PCCT and 17.4% for CBCT. For horizontal defect width (mean reference 0.99 ± 0.39 mm), both modalities slightly underestimated the defect width compared to the reference standard of the oral scanner. However, PCCT showed lower MAE (0.077 mm vs. 0.18 mm; p < 0.001) and reduced proportional variability. Intra- and inter-reader reliability were consistently higher for PCCT (ICC up to 0.994) compared with CBCT.Within the limitations of this ex vivo model, PCCT provided more accurate and reproducible three-dimensional measurements of standardized peri-implant bone defects than CBCT. These findings need to be confirmed by further in vivo studies to determine its clinical diagnostic value and potential impact on treatment planning.

Keyword(s): Animals (MeSH) ; Peri-Implantitis: diagnostic imaging (MeSH) ; Swine (MeSH) ; Cone-Beam Computed Tomography (MeSH) ; Tomography, X-Ray Computed: methods (MeSH) ; Photons (MeSH) ; Dental Implants (MeSH) ; Mandible: diagnostic imaging (MeSH) ; Reproducibility of Results (MeSH) ; 3D dental imaging ; CBCT ; Image quality ; Implant diagnostics ; Peri-implantitis ; Photon-counting CT ; Dental Implants

Classification:

Note: #LA:E025#

Contributing Institute(s):
  1. Radiologie (E010)
  2. Röntgenbildgebung und Computertomographie (E025)
Research Program(s):
  1. 315 - Bildgebung und Radioonkologie (POF4-315) (POF4-315)

Appears in the scientific report 2026
Database coverage:
Medline ; Creative Commons Attribution CC BY (No Version) ; DOAJ ; Article Processing Charges ; Clarivate Analytics Master Journal List ; Current Contents - Clinical Medicine ; DOAJ Seal ; Essential Science Indicators ; Fees ; IF < 5 ; JCR ; PubMed Central ; Science Citation Index Expanded ; Web of Science Core Collection
Click to display QR Code for this record

The record appears in these collections:
Document types > Articles > Journal Article
Institute Collections > E010
Institute Collections > E025
Public records
Publications database

 Record created 2026-11-04, last modified 2026-11-04


Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)