Prompt gamma spectroscopy for absolute range verification of 12C ions at synchrotron based facilities.

 ;  ;  ;  ;  ;  ;

2020
IOP Publ. Bristol

Physics in medicine and biology 65(9), 095010 () [10.1088/1361-6560/ab7973]
 GO

This record in other databases:

Please use a persistent id in citations: doi:

Abstract: The physical range uncertainty limits the exploitation of the full potential of charged particle therapy. In this work we face this issue aiming to measure the absolute Bragg peak position in the target. We investigate p,4He,12C and16O beams accelerated at the Heidelberg Ion-Beam Therapy Center. The residual range of the primary12C ions is measured using the prompt gamma spectroscopy method, which was demonstrated for proton beams accelerated by cyclotrons and is developed here for the first time for heavier ions accelerated by a synchrotron. We develop a detector system that includes (i) a spectroscopic unit based on cerium(III) bromide and bismuth germanium oxide scintillating crystals, (ii) a beam trigger based on an array of scintillating fibers and (iii) a data acquisition system based on a FlashADC. We test the system in two different scenarios. In the first series of experiments we detect and identify 19 independent spectral lines over a wide gamma energy spectrum in presence of the four ion species for different targets, including also a water target with a titanium insert. In the second series of experiments we introduce a collimator aiming to relate the spectral information to the range of the primary particles. We perform extensive measurements for a12C beam and demonstrate submillimetric precision for the measurement of its Bragg peak position in the experimental setup. The features of the energy and time spectra for gamma radiation induced by p,4He and16O are investigated upstream and downstream the Bragg peak position. We conclude the analysis extrapolating the required future developments, which would be needed to achieve range verification with a 2 mm accuracy during a single fraction delivery of D = 2 Gy physical dose.

Classification:

Note: 2020 May 11;65(9):095010#EA:E041#LA:E041#

Contributing Institute(s):
  1. E041 Medizinische Physik in der Radioonkologie (E041)
Research Program(s):
  1. 315 - Imaging and radiooncology (POF3-315) (POF3-315)

Appears in the scientific report 2020
Database coverage:
Medline ; BIOSIS Previews ; Clarivate Analytics Master Journal List ; Current Contents - Life Sciences ; Ebsco Academic Search ; IF < 5 ; JCR ; National-Konsortium ; NationallizenzNationallizenz ; SCOPUS ; Science Citation Index ; Science Citation Index Expanded ; Web of Science Core Collection
Click to display QR Code for this record

The record appears in these collections:
Document types > Articles > Journal Article
Public records
Publications database

 Record created 2020-02-25, last modified 2024-02-29



Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)